Optical frequency domain imaging with a rapidly swept laser in the 815-870 nm range.
???displayArticle.abstract???
Optical frequency domain imaging (OFDI) in the 800-nm biological imaging window is demonstrated by using a novel wavelength-swept laser source. The laser output is tuned continuously from 815 to 870 nm at a 43.2-kHz repetition rate with 7-mW average power. Axial resolution of 10-mum in biological tissue and peak sensitivity of 96 dB are achieved. In vivo imaging of Xenopus laevis is demonstrated with an acquisition speed of 84 frames per second (512 axial lines per frame). This new imaging technique may prove useful in comprehensive retinal screening for medical diagnosis and contrast-agent-based imaging for biological investigations.
???displayArticle.pubmedLink??? 19516763
???displayArticle.link??? Opt Express
Species referenced: Xenopus laevis